Cookies on this website

We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND:We sought to compare the early hemodynamic performance of pericardial stentless aortic valves with that of well-established porcine stentless aortic prostheses. METHODS:A total of 169 patients (97 men and 72 women, aged 73+/-6 years) undergoing aortic valve replacement received either a pericardial (Pericarbon, Sorin Biomedica, Saluggia, Italy; n = 89) or a porcine (Freestyle, Medtronic, n = 80) stentless aortic valve. Aortic valve hemodynamics and root dynamism were assessed by Doppler echocardiography at discharge and 12 months after implantation. RESULTS:Clinical demographic data, valve size (24.0+/-1.9 vs 24.6+/-2.3 mm), and body surface area (1.85+/-0.19 vs 1.80+/-0.19 m2) did not differ between porcine and pericardial valves (both p > 0.05). The 1-year postoperative mean valve pressure gradient (4.2+/-2.6 vs 3.7+/-2.6 mm Hg), effective orifice area (2.2+/-0.8 vs 2.2+/-0.8 cm2), and left ventricular ejection fraction (62+/-13 vs 63+/-13, %) also did not differ (all p > 0.05). However, at discharge, systolic increase in aortic sinus diameter was significantly greater in pericardial valves than in porcine ones (7.7+/-5.7 vs 4.9%+/-4.2%, p < 0.01). Furthermore, pericardial valves had a greater slope of effective orifice area-systolic aortic flow relationship (0.89+/-0.07 vs 0.70+/-0.06, cm2/100 mL/s, p < 0.01). CONCLUSIONS:Nonprosthetic thin-walled pericardial valves appear to offer better aortic root dynamism and more efficient hemodynamics than those of porcine valves immediately after implant. At 1-year follow-up, however, both types of stentless valves provide equally excellent hemodynamics. The clinical choice between the two will depend on their long-term durability.

Original publication

DOI

10.1016/s0003-4975(01)02553-x

Type

Journal article

Journal

The Annals of thoracic surgery

Publication Date

05/2001

Volume

71

Pages

S311 - S314

Addresses

Department of Cardiac Surgery, Oxford Heart Centre, Oxford Radcliffe Hospitals, United Kingdom. x.y.jin@btinternet.com

Keywords

Aortic Valve, Humans, Postoperative Complications, Prosthesis Failure, Echocardiography, Doppler, Survival Rate, Follow-Up Studies, Prospective Studies, Prosthesis Design, Bioprosthesis, Heart Valve Prosthesis, Stents, Aged, Aged, 80 and over, Middle Aged, Female, Male, Hemodynamics