We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Wireless left ventricular endocardial pacing with the WiSE CRT system has recently become available as alternative to conventional epicardial CRT pacing. We report the first comparison of the acute electrical and hemodynamic response produced by the two CRT pacing modalities in a patient undergoing WiSE CRT implant after a failed conventional CRT procedure. WiSE CRT pacing showed an additive acute benefit compared with conventional CRT. These findings could potentially translate into long-term clinical benefit and introduce the potential for tri-ventricular pacing using both systems simultaneously.

Original publication

DOI

10.1002/joa3.12019

Type

Journal article

Journal

J Arrhythm

Publication Date

02/2018

Volume

34

Pages

87 - 89

Keywords