We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Mass spectrometry has recently been proposed as a novel sensitive screening tool for monoclonal gammopathies. In a small study we have tested the ability of quantitative immunoprecipitation mass spectrometry (QIP-MS) to identify low level monoclonal immunoglobulins not currently detected by the initial serum protein electrophoresis (SPEP) screen. QIP-MS positively identified the primary monoclonal immunoglobulins in all 11 patient samples alongside additional monoclonal immunoglobulins in a subset of patient samples. We conclude that QIP-MS has clinical utility as a first-line screening tool for monoclonal gammopathy investigation, identifying monoclonality in patients with higher sensitivity and resolution compared to the current standard methods.

Original publication

DOI

10.1016/j.clinbiochem.2021.05.012

Type

Journal article

Journal

Clin Biochem

Publication Date

09/2021

Volume

95

Pages

81 - 83

Keywords

Mass spectrometry, Monoclonal gammopathy, Monoclonal protein, Multiple myeloma