We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

A new method is proposed for reducing the artifacts produced in diffusion-weighted imaging. When data are acquired using multiple shot echo planar acquisitions, conventional reconstruction methods produce artifactual images as a consequence of diffusion weighting and small amounts of bulk motion of the subject. If the amount of motion can be determined, it is possible to correct the data before reconstruction, which removes the artifact. A method for estimating the motion from the acquired data has been developed and evaluated. This method assumes that ghost image effects will be minimized when motion has been correctly compensated. By considering the amount of signal in the background of the image, appropriate corrections to the data can be made, and the accuracy of the motion compensation may be estimated. This technique has been evaluated by computer simulation, and its performance has been demonstrated in a phantom and humans with both two- and four-shot echo planar acquisitions and using both "mosaic" and "interleaved" sampling schemes.

Type

Journal article

Journal

Magn Reson Med

Publication Date

07/1997

Volume

38

Pages

82 - 88

Keywords

Artifacts, Brain, Computer Simulation, Echo-Planar Imaging, Humans, Phantoms, Imaging