We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVE: To report a first case series of the Lotus valve (Boston Scientific) for the treatment of surgical aortic bioprosthesis failure. BACKGROUND: Valve-in-valve (VIV) transcatheter aortic valve implantation (TAVI) is an established management option for surgical aortic bioprosthesis failure. Although TAVI has proven efficacious, complications relating to suboptimal valve implantation have been reported in approximately 5% of patients. The Lotus valve is fully repositionable and retrievable, and therefore enables complete assessment of valve function prior to definitive deployment. However, data supporting its use for this indication are limited. METHODS: Consecutive patients who underwent transfemoral VIV-TAVI at John Radcliffe Hospital between January 2015 and December 2016 were retrospectively analyzed. Procedural and 30-day outcomes were reported in accordance with Valve Academic Research Consortium-2 (VARC-2) definitions. RESULTS: Seven patients underwent VIV-TAVI with the Lotus valve. Device success according to VARC-2 criteria was achieved in 6 of the 7 patients. Following implantation, transvalvular hemodynamics were acceptable, with a mean gradient of 11.9 ± 6.6 mm Hg. All patients had mild or no residual aortic regurgitation. Specifically, no patient required further valve-related intervention, or suffered myocardial infarction, stroke, or acute kidney injury stage 2 or 3. There were no deaths during the follow-up period. CONCLUSIONS: Transfemoral VIV-TAVI for the treatment of surgical aortic bioprosthesis failure with the Lotus device appears to be safe and is associated with no significant residual aortic regurgitation, and offers favorable transvalvular hemodynamics at 30-day follow-up.

Type

Journal article

Journal

J Invasive Cardiol

Publication Date

09/2017

Volume

29

Pages

315 - 319

Keywords

Aged, Aged, 80 and over, Aortic Valve, Aortic Valve Insufficiency, Bioprosthesis, Female, Heart Valve Prosthesis, Humans, Male, Prosthesis Design, Reoperation, Retrospective Studies, Transcatheter Aortic Valve Replacement