We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

(23)Na MRI may allow distinction of normal and ischemically injured myocardium. The aim of this study was to optimize (23)Na MRI of the human heart by improvement of spatial resolution and ECG-triggering and to measure the signal/noise of blood and myocardium and the myocardium/blood signal ratios in a volunteer study. A spoiled gradient echo sequence was developed on a 1.5T scanner equipped with a (23)Na heart surface coil. 3D short axis ECG-triggered (23)Na MRI was performed in 10 healthy subjects. The signal/noise of myocardium and blood were 8.2 +/- 0.7 and 18.3 +/- 1.3, respectively, signal ratio myocardium/blood was 0.44 +/- 0.03. This value is in good agreement with the theoretical value of 0.45. ECG gated 3D (23)Na MRI of the human heart is feasible with sufficient spatial resolution and signal/noise ratio. Magn Reson Med 45:164-166, 2001.

Type

Journal article

Journal

Magn Reson Med

Publication Date

01/2001

Volume

45

Pages

164 - 166

Keywords

Adult, Electrocardiography, Female, Heart, Humans, Image Enhancement, Magnetic Resonance Imaging, Male, Sodium